INSIGNIA
Report
- Report Number
- 2124215-2014-08451
- Event Type
- Malfunction
- Date Received
- July 1, 2014
- Date of Event
- November 2, 2007
- Report Date
- April 11, 2014
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- LWP
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
(B)(4). UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, THIS DEVICE WAS THOROUGHLY INSPECTED AND ANALYZED. THE DEVICE WAS EXPOSED TO SIMULATED HEART LOAD CONDITIONS, AND THE PACING AND SENSING FUNCTIONS WERE TESTED. THE DEVICE OPERATED APPROPRIATELY, ACCORDING TO ITS PERFORMANCE SPECIFICATIONS, WITH NO INTERRUPTIONS IN THERAPY OUTPUT OR PROGRAMMER COMMUNICATION AT THE RETURNED PROGRAMMED SETTINGS. IT WAS NOTED THAT THE DEVICE BATTERY MONITORING VOLTAGE WAS 2.69 VOLTS. THE DEVICE CASE WAS OPENED TO FACILITATE EXAMINATION OF THE INTERNAL COMPONENTS. THE DEVICE BATTERY WAS REMOVED AND REPLACED WITH AN EXTERNAL POWER SOURCE. DETAILED ANALYSIS OF THE DEVICE HYBRID WAS UNDERTAKEN. ANALYSIS DID NOT IDENTIFY ANY CONDITIONS THAT WOULD HAVE CAUSED OR CONTRIBUTED TO THE IDENTIFIED FAILURE OF THE DEVICE TO MEET THE LABELED LONGEVITY CALCULATION. THE ROOT CAUSE COULD NOT BE IDENTIFIED.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT A COMPETITOR RIGHT VENTRICULAR (RV) LEAD CAUSED A LEAD SAFETY SWITCH TO OCCUR WITH THIS PACEMAKER. A TECHNICAL SERVICE CONSULTANT DISCUSSED THE REASONS THAT TRIGGER MODE SWITCHING AND INDICATED THAT IT IS NECESSARY TO CHECK DAILY MEASUREMENTS TO VERIFY ELECTRICAL LEAD MEASUREMENTS AND PERFORM A NEW PACING IMPEDANCE AND THRESHOLD TESTING TO VERIFY LEAD INTEGRITY. THE PACEMAKER WAS EXPLANTED SEVERAL YEARS LATER WITH NO ADVERSE PATIENT EFFECTS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 385395 | INSIGNIA | IMPLANTABLE PULSE GENERATOR | LWP | GUIDANT CRM CLONMEL IRELAND | 1291 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 1291 |