FDA Adverse Event Malfunction Summary report: N

INSIGNIA

MDR report key: 3904160 · Received July 1, 2014

Report

Report Number
2124215-2014-09869
Event Type
Malfunction
Date Received
July 1, 2014
Date of Event
April 1, 2014
Report Date
June 8, 2014
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWP
PMA / PMN Number
D970003
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IA
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE IS EXPECTED TO BE RETURNED FOR TESTING. THIS PRODUCT ISSUE WILL BE UPDATED WHEN THE DEVICE IS RETURNED AND EVALUATION IS COMPLETE.

Additional Manufacturer Narrative · 1

(B)(4). UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE DEVICE WAS PERFORMED. A VISUAL INSPECTION OF THE DEVICE HEADER AND CASE NOTED NO ANOMALIES. THE DEVICE WAS THEN EXPOSED TO SIMULATED HEART LOAD CONDITIONS, AND THE PACING, AND SENSING FUNCTIONS WERE TESTED. IMPEDANCE TESTING WAS COMPLETED AND ALL MEASUREMENTS WERE WITHIN NORMAL LIMITS. THE DEVICE OPERATED APPROPRIATELY WITH NO INTERRUPTIONS IN THERAPY OUTPUT AT THE RETURNED PROGRAMMED SETTINGS. A SERIES OF ELECTRICAL TESTS WAS ALSO PERFORMED, AND AGAIN, NORMAL DEVICE FUNCTION WAS OBSERVED.

Description of Event or Problem · 1

--

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS SYSTEM HAD BEEN EXHIBITING GRADUALLY INCREASING RIGHT VENTRICULAR (RV) IMPEDANCE AND THRESHOLDS MEASUREMENTS. THE DEVICE WAS ELECTIVELY REPLACED DUE TO NORMAL BATTERY DEPLETION. DURING THE PROCEDURE, ONE OF THE SETSCREWS WAS LOOSE AND IT WAS SUSPECTED THAT IT HAD NOT BEEN MAKING ADEQUATE CONNECTION WITH THE CHRONIC LEAD. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
383364 INSIGNIA IMPLANTABLE PULSE GENERATOR LWP GUIDANT CRM CLONMEL IRELAND 1298

Patients

Seq Age Sex Outcome Treatment
1 84 YR 4086| 4087| 4136| K063| 1298