FDA Adverse Event Malfunction Summary report: N

NON-HEATED ADULT BREATHING CIRCUIT

MDR report key: 3903952 · Received June 30, 2014

Report

Report Number
9611451-2014-00581
Event Type
Malfunction
Date Received
June 30, 2014
Date of Event
June 2, 2014
Report Date
June 10, 2014
Manufacturer
FISHER & PAYKEL HEALTHCARE LTD
Product Code
BZO
PMA / PMN Number
SEE H10
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). METHOD: THE RETURNED COMPLAINT RT134 BREATHING CIRCUIT WAS VISUALLY INSPECTED, PRESSURE TESTED AND IMMERSED IN A WATER BATH TO CHECK FOR LEAK. RESULTS: PRESSURE TESTING REVEALED THAT THE CIRCUIT WAS OUT OF SPECIFICATION. THE WATER BATH TEST REVEALED THAT THE LEAK WAS IN THE EXPIRATORY WATER TRAP. A LOT CHECK REVEALED NO OTHER COMPLAINTS FOR LOT NUMBER 130923. CONCLUSION: THE WATER TRAP OF THE BREATHING CIRCUIT CONSISTS OF TWO PARTS, A LID AND A BOWL, WHICH CAN BE DISCONNECTED DURING USE FOR DRAINING WATER. THE WATER TRAP LEAK WAS LOCATED AT THE SEAL BETWEEN THE WATER TRAP BOWL AND THE WATER TRAP LID. ALL CIRCUITS ARE PRESSURE TESTED FOR LEAKS DURING PRODUCTION AND THOSE THAT FAIL ARE REJECTED. THIS SUGGESTS ANY LEAK MUST HAVE DEVELOPED POST PRODUCTION DURING TRANSPORT, STORAGE OR USE, POSSIBLY BY DISTORTION OF THE WATER TRAP WHEN THE BOWL WAS CONNECTED. THE USER INSTRUCTIONS THAT ACCOMPANY THE RT134 ADULT BREATHING CIRCUIT STATE THE FOLLOWING: PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT. SET APPROPRIATE VENTILATOR ALARMS. HOSPITAL STAFF CORRECTLY CHECKED THE BREATHING CIRCUIT BEFORE PATIENT USE, WHICH IS IN LINE WITH OUR USER INSTRUCTIONS.

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED TO A FISHER & PAYKEL HEALTHCARE FIELD REPRESENTATIVE THAT AN RT134 ADULT BREATHING CIRCUIT FAILED THE SERVO-I VENTILATOR LEAK TEST. THIS WAS FOUND BEFORE USE ON A PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
379825 NON-HEATED ADULT BREATHING CIRCUIT BZO BZO FISHER & PAYKEL HEALTHCARE LTD RT134 130923

Patients

Seq Age Sex Outcome Treatment
1 SERVO-I VENTILATOR