ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-18707
- Event Type
- Injury
- Date Received
- June 30, 2014
- Report Date
- June 24, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 08/14/2014 WITH THE FOLLOWING FINDINGS: EVALUATION REVEALED THAT THE LAST BASAL DELIVERY WAS ON (B)(4) 2014 AT 12:50PM. THE BLACK BOX SHOWS SEVERAL ¿NO CARTRIDGE DETECTED¿ WARNINGS ON (B)(4) 2014. TOTAL DAILY DOSE ADDS UP AND REFLECTS THE PROGRAMMED BASAL RATE TARGET. THE DISPLAY IS DIM AND FADED; THE PISTON WAS UNABLE TO RECOGNIZE THE CARTRIDGE UPON THE FIRST LOAD ATTEMPT. THE COVER WAS REMOVED AND MOISTURE CORROSION WAS FOUND TO THE PCB ON THE U2 COMPONENT, THE FORCE SENSOR FLEX PINS AND FORCE SENSOR PLATE. FORCE SENSOR RESISTANCE IS WITHIN SPECIFICATIONS AT 9.4K OHMS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
THE PATIENT CONTACTED ANIMAS ON (B)(6) 2014 ALLEGING PRIME (LOAD STEP MALFUNCTION) ISSUE. THE PATIENT REPORTED THAT THEIR BLOOD GLUCOSE (BG) WAS 38MG/DL AND WAS UNABLE TO GET BG ABOVE 60MG/DL FOR A PERIOD OF ABOUT FOUR TO FIVE HOURS. THE PATIENT STATED THAT THEY WERE UNSURE HOW MUCH WAS INFUSED INTO THEM. THE PATIENT STATED THAT THERE WAS A FILLED CARTRIDGE IN PUMP. THE PATIENT CONFIRMED THAT THE PUMP WAS PRIMING TUBING DURING LOAD STEP. THIS REPORT IS BEING MADE DUE TO BG EXCURSION, THE PATIENT EXPERIENCED DUE TO AN ALLEGED LOAD STEP MALFUNCTION ISSUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 380730 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 32 YR | Life Threatening |