ASR ACETABULAR CUPS 58
Report
- Report Number
- 1818910-2014-22331
- Event Type
- Injury
- Date Received
- June 30, 2014
- Date of Event
- April 30, 2014
- Report Date
- June 27, 2014
- Manufacturer
- DEPUY INTERNATIONAL LTD.
- Product Code
- KWA
- PMA / PMN Number
- PK040627
- Removal / Correction Number
- Z-1749/1816-2011
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- PATIENT
Narratives
ADDITIONAL NARRATIVE:IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.THE ASR PLATFORM WAS VOLUNTARILY RECALLED FROM THE MARKET IN AUGUST 2010, AND THE ASR PRODUCT CODES ARE NOW CONSIDERED INACTIVE. FURTHER INVESTIGATION OF THIS INDIVIDUAL INCIDENT WILL NOT BE UNDERTAKEN, AS THERE IS AN ONGOING INVESTIGATION REGARDING THE ROOT CAUSE(S) AND/OR CORRECTIVE ACTIONS. (B)(4). DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD THE PRODUCT AND/OR ADDITIONAL INFORMATION BE RECEIVED, THE INVESTIGATION WILL BE RE-OPENED.
IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.(B)(4). DEPUY STILL CONSIDERS THIS INVESTIGATION CLOSED.
PATIENT CONTACTED DEPUY AS A RESULT OF THE ASR RECALL TO INITIATE A CLAIM. MEDICAL RECORDS WERE OBTAINED. MEDICAL RECORDS INDICATE PATIENT WAS REVISED DUE TO PAIN, STEM LOOSENING, ACETABULAR CUP LOOSENING AND PSEUDOTUMOR FROM METAL ON METAL HIP ARTHROPLASTY. UPON REVISION, CORROSION AT THE FEMORAL HEAD/TRUNNION JUNCTION AND A SMALL CRACK IN THE TROCHANTER WERE NOTED.
UPDATE REC'D 6/27/2014 - LITIGATION RECEIVED. LITIGATION ALLEGES STIFFNESS, DISCOMFORT, WEAKNESS AND ELEVATED METAL ION LEVELS. THERE IS NO NEW ADDITIONAL INFORMATION THAT WOULD AFFECT THE INVESTIGATION. THIS COMPLAINT WAS UPDATED ON: 07/10/2014.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 380996 | ASR ACETABULAR CUPS 58 | ACETABULAR CUP | KWA | DEPUY INTERNATIONAL LTD. | 2112731 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 77 YR | Required Intervention |