FDA Adverse Event Injury Summary report: N

ASR ACETABULAR CUPS 58

MDR report key: 3903829 · Received June 30, 2014

Report

Report Number
1818910-2014-22331
Event Type
Injury
Date Received
June 30, 2014
Date of Event
April 30, 2014
Report Date
June 27, 2014
Manufacturer
DEPUY INTERNATIONAL LTD.
Product Code
KWA
PMA / PMN Number
PK040627
Removal / Correction Number
Z-1749/1816-2011
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

ADDITIONAL NARRATIVE:IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.THE ASR PLATFORM WAS VOLUNTARILY RECALLED FROM THE MARKET IN AUGUST 2010, AND THE ASR PRODUCT CODES ARE NOW CONSIDERED INACTIVE. FURTHER INVESTIGATION OF THIS INDIVIDUAL INCIDENT WILL NOT BE UNDERTAKEN, AS THERE IS AN ONGOING INVESTIGATION REGARDING THE ROOT CAUSE(S) AND/OR CORRECTIVE ACTIONS. (B)(4). DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD THE PRODUCT AND/OR ADDITIONAL INFORMATION BE RECEIVED, THE INVESTIGATION WILL BE RE-OPENED.

Additional Manufacturer Narrative · 1

IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.(B)(4). DEPUY STILL CONSIDERS THIS INVESTIGATION CLOSED.

Description of Event or Problem · 1

PATIENT CONTACTED DEPUY AS A RESULT OF THE ASR RECALL TO INITIATE A CLAIM. MEDICAL RECORDS WERE OBTAINED. MEDICAL RECORDS INDICATE PATIENT WAS REVISED DUE TO PAIN, STEM LOOSENING, ACETABULAR CUP LOOSENING AND PSEUDOTUMOR FROM METAL ON METAL HIP ARTHROPLASTY. UPON REVISION, CORROSION AT THE FEMORAL HEAD/TRUNNION JUNCTION AND A SMALL CRACK IN THE TROCHANTER WERE NOTED.

Description of Event or Problem · 1

UPDATE REC'D 6/27/2014 - LITIGATION RECEIVED. LITIGATION ALLEGES STIFFNESS, DISCOMFORT, WEAKNESS AND ELEVATED METAL ION LEVELS. THERE IS NO NEW ADDITIONAL INFORMATION THAT WOULD AFFECT THE INVESTIGATION. THIS COMPLAINT WAS UPDATED ON: 07/10/2014.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
380996 ASR ACETABULAR CUPS 58 ACETABULAR CUP KWA DEPUY INTERNATIONAL LTD. 2112731

Patients

Seq Age Sex Outcome Treatment
1 77 YR Required Intervention