FDA Adverse Event
Malfunction
Summary report: N
CORE IMPACTION DRILL
MDR report key: 3903807
·
Received June 30, 2014
Report
- Report Number
- 0001811755-2014-02337
- Event Type
- Malfunction
- Date Received
- June 30, 2014
- Date of Event
- June 5, 2014
- Report Date
- June 5, 2014
- Manufacturer
- STRYKER INSTRUMENTS-KALAMAZOO
- Product Code
- DZJ
- PMA / PMN Number
- K040369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE TECHNICIAN WAS ABLE TO DUPLICATE THE REPORTED EVENT AND NOTED THAT THE MOTOR ASSEMBLY WAS CORRODED.
Description of Event or Problem · 1
IT WAS REPORTED BY THE USER FACILITY THAT THE CORE IMPACTION DRILL WAS OVERHEATING. THE USER FACILITY WAS NOT ABLE TO PROVIDE ANY FURTHER INFORMATION REGARDING THE REPORTED EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 380947 | CORE IMPACTION DRILL | DRIVER, WIRE, AND BONE DRILL, MANUAL | DZJ | STRYKER INSTRUMENTS-KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |