FDA Adverse Event Injury Summary report: N

RIATA ST PASSIVE FIXATION

MDR report key: 3903748 · Received June 30, 2014

Report

Report Number
2938836-2014-12750
Event Type
Injury
Date Received
June 30, 2014
Date of Event
April 21, 2014
Manufacturer
ST. JUDE MEDICAL, INC., CRMD
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
Z0458
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. EVALUATION DESCRIPTION INCLUDED. EXTERNAL INSULATION ABRASIONS WERE NOTED AT 7.6-8.1CM AND 7.7-7.9CM FROM THE DISTAL TIP, CONSISTENT WITH LEAD FRICTION TO ANOTHER DEVICE OR FEATURE OF THE HEART. THE ETFE COATING WAS INTACT AT THESE LOCATIONS. EXTERNAL INSULATION ABRASION WAS NOTED AT 7.4-8.1CM FROM THE DISTAL TIP, CONSISTENT WITH LEAD FRICTION TO ANOTHER DEVICE OR FEATURE OF THE HEART. THE ETFE COATING WAS ABRADED AT THIS LOCATION.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE ASYMPTOMATIC PATIENT PRESENTED IN THE OPERATING ROOM FOR A DEVICE UPGRADE. THE LEAD WAS EXPLANTED AND REPLACED. EXTERNALIZED CONDUCTORS WERE OBSERVED. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
379960 RIATA ST PASSIVE FIXATION DEFIBRILLATION LEAD NVY ST. JUDE MEDICAL, INC., CRMD 7040/65 NA

Patients

Seq Age Sex Outcome Treatment
1 59 YR