FDA Adverse Event
Injury
Summary report: N
RIATA ST PASSIVE FIXATION
MDR report key: 3903748
·
Received June 30, 2014
Report
- Report Number
- 2938836-2014-12750
- Event Type
- Injury
- Date Received
- June 30, 2014
- Date of Event
- April 21, 2014
- Manufacturer
- ST. JUDE MEDICAL, INC., CRMD
- Product Code
- NVY
- PMA / PMN Number
- P950022
- Removal / Correction Number
- Z0458
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KY, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. EVALUATION DESCRIPTION INCLUDED. EXTERNAL INSULATION ABRASIONS WERE NOTED AT 7.6-8.1CM AND 7.7-7.9CM FROM THE DISTAL TIP, CONSISTENT WITH LEAD FRICTION TO ANOTHER DEVICE OR FEATURE OF THE HEART. THE ETFE COATING WAS INTACT AT THESE LOCATIONS. EXTERNAL INSULATION ABRASION WAS NOTED AT 7.4-8.1CM FROM THE DISTAL TIP, CONSISTENT WITH LEAD FRICTION TO ANOTHER DEVICE OR FEATURE OF THE HEART. THE ETFE COATING WAS ABRADED AT THIS LOCATION.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE ASYMPTOMATIC PATIENT PRESENTED IN THE OPERATING ROOM FOR A DEVICE UPGRADE. THE LEAD WAS EXPLANTED AND REPLACED. EXTERNALIZED CONDUCTORS WERE OBSERVED. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 379960 | RIATA ST PASSIVE FIXATION | DEFIBRILLATION LEAD | NVY | ST. JUDE MEDICAL, INC., CRMD | 7040/65 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 59 YR |