FDA Adverse Event Malfunction Summary report: N

UNIFY ASSURA DR CRT-D, DF-4 CONNECTOR

MDR report key: 3903742 · Received June 30, 2014

Report

Report Number
2938836-2014-12746
Event Type
Malfunction
Date Received
June 30, 2014
Date of Event
May 26, 2014
Manufacturer
ST. JUDE MEDICAL, INC., CRMD
Product Code
NIK
PMA / PMN Number
P030054
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED THAT AN ASYMPTOMATIC PATIENT PRESENTED FOR A FOLLOW-UP. EPISODES OF NON-SUSTAINED RV OVERSENSING WERE OBSERVED. THE DEVICE WAS INTERMITTENTLY OVERSENSING. PROGRAMMING CHANGES WERE MADE AND RESOLVED THE OVERSENSING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
379958 UNIFY ASSURA DR CRT-D, DF-4 CONNECTOR IMPLANTABLE CARDIOVERTER DEFIBRILLATOR NIK ST. JUDE MEDICAL, INC., CRMD CD3357-40Q NA

Patients

Seq Age Sex Outcome Treatment
1 59 YR