FDA Adverse Event Malfunction Summary report: N

QUADRA ASSURA CRT-D, DF-4 CONNECTOR

MDR report key: 3903723 · Received June 30, 2014

Report

Report Number
2938836-2014-12886
Event Type
Malfunction
Date Received
June 30, 2014
Date of Event
May 30, 2014
Manufacturer
ST. JUDE MEDICAL, INC., CRMD
Product Code
NIK
PMA / PMN Number
P910023
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED THAT AN ASYMPTOMATIC PATIENT PRESENTED IN CLINIC FOR FOLLOW UP. THE DEVICE EXHIBITED POST-PACED T-WAVE OVERSENSING. PROGRAMMING CHANGES WERE RECOMMENDED. NO FURTHER INFORMATION IS AVAILABLE.

Description of Event or Problem · 1

NEW INFORMATION NOTES THAT THE DEVICE WAS REPROGRAMMED AND THE OVERSENSING WAS RESOLVED. THE PATIENT WAS FINE AFTER THE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
380317 QUADRA ASSURA CRT-D, DF-4 CONNECTOR IMPLANTABLE CARDIOVERTER DEFIBRILLATOR NIK ST. JUDE MEDICAL, INC., CRMD CD3365-40Q NA

Patients

Seq Age Sex Outcome Treatment
1 88 YR