FDA Adverse Event
Malfunction
Summary report: N
UNIFY ASSURA DR CRT-D
MDR report key: 3903624
·
Received June 30, 2014
Report
- Report Number
- 2938836-2014-12759
- Event Type
- Malfunction
- Date Received
- June 30, 2014
- Date of Event
- April 28, 2014
- Manufacturer
- ST. JUDE MEDICAL, INC., CRMD
- Product Code
- NIK
- PMA / PMN Number
- P030054
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AL, US
- Reporter Occupation
- NURSE
Narratives
Description of Event or Problem · 1
IT WAS REPORTED THAT VIA REMOTE TRANSMISSION, NON-SUSTAINED LEAD NOISE DUE TO POST-PACED T-WAVE OVERSENSING WAS OBSERVED. PATIENT WAS ASYMPTOMATIC. PROGRAMMING CHANGES WERE RECOMMENDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 381106 | UNIFY ASSURA DR CRT-D | IMPLANTABLE CARDIOVERTER DEFIBRILLATOR | NIK | ST. JUDE MEDICAL, INC., CRMD | CD3357-40C | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 80 YR |