FDA Adverse Event Malfunction Summary report: N

COULTER LH 750 HEMATOLOGY ANALYZER

MDR report key: 3903612 · Received June 30, 2014

Report

Report Number
1061932-2014-01501
Event Type
Malfunction
Date Received
June 30, 2014
Date of Event
June 4, 2014
Report Date
June 4, 2014
Manufacturer
BECKMAN COULTER
Product Code
GKZ
PMA / PMN Number
K011342
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A BECKMAN COULTER (BEC) FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO ASSESS THE INSTRUMENT. THE FSE OBSERVED THE PINCH VALVE VL57A AT FITTING 68 WAS COMING OFF. THE FSE REPLACED THE PINCH VALVE VL57A, WHICH REPAIRED THE LEAK. THE FSE VERIFIED THE INSTRUMENT AND THE INSTRUMENT WAS RETURNED TO NORMAL OPERATION. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER REPORTED A LEAK FROM THE NEEDLE BELLOWS INSIDE THE COULTER LH 750 HEMATOLOGY ANALYZER. THE VOLUME OF THE LEAK IS UNKNOWN. THE LEAK WAS CONTAINED WITHIN THE INSTRUMENT. THE CUSTOMER WAS WEARING PERSONAL PROTECTION EQUIPMENT (PPE) CONSISTING OF GLOVES, FACE SHIELD AND A LAB COAT AT THE TIME OF THE OCCURRENCE. THERE WAS NO REPORTED EXPOSURE TO MUCOUS MEMBRANES OR OPEN WOUNDS. NO ERRONEOUS RESULTS WERE GENERATED IN CONNECTION WITH THE REPORTED EVENT. THERE WAS NO DEATH, INJURY, OR AFFECT TO USER OR TO PATIENT TREATMENT ATTRIBUTED TO THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
379561 COULTER LH 750 HEMATOLOGY ANALYZER COUNTER, DIFFERENTIAL CELL GKZ BECKMAN COULTER NA NA

Patients

Seq Age Sex Outcome Treatment
1