FDA Adverse Event Malfunction Summary report: N

HOMECHOICE

MDR report key: 3903546 · Received June 30, 2014

Report

Report Number
1416980-2014-20981
Event Type
Malfunction
Date Received
June 30, 2014
Date of Event
June 5, 2014
Report Date
June 5, 2014
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE HOMECHOICE (HC) DEVICE WAS RETURNED AND EVALUATED BY THE PRODUCT ANALYSIS LAB (PAL). THE EVENT HISTORY LOG REVIEW SHOWED NO KEYSTROKES, PROGRAMMING, OR USE RELATED EVENTS THAT INDICATED AND/OR CONTRIBUTED TO THE RITE (RETURNED INSTRUMENT TEST EVALUATION) FAILURE. THE EXTERNAL/INTERNAL INSPECTION WAS PERFORMED AND THE DEVICE PASSED, ALONG WITH ALL OF THE ELECTRICAL TESTING. DURING THE EVALUATION, THE FLUID WAS TRANSFERRED OUTSIDE OF THE RITE SPECIFIED LIMITS; THE VOLUMETRIC ACCURACY TEST THAT FOLLOWED SHOWED THAT THE DEVICE HAD FAILED. THE RESULTS OF THE EVALUATION REVEALED THE CAUSE OF THE FAILURE TO BE DETERIORATED PISTON FOAM. THE PISTON FOAM WAS TO BE SCRAPPED AND THE DEVICE WAS SENT FOR SERVICING. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE HAS BEEN RECEIVED AND THE EVALUATION IS IN PROGRESS. UPON COMPLETION OF THE DEVICE EVALUATION, OR IF ANY ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED HOMECHOICE DEVICE, THE DEVICE FAILED FLUID VOLUME ACCURACY TESTING. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
379174 HOMECHOICE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment
1