HOMECHOICE
Report
- Report Number
- 1416980-2014-20981
- Event Type
- Malfunction
- Date Received
- June 30, 2014
- Date of Event
- June 5, 2014
- Report Date
- June 5, 2014
- Manufacturer
- BAXTER HEALTHCARE - LARGO
- Product Code
- FKX
- PMA / PMN Number
- K102936
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE HOMECHOICE (HC) DEVICE WAS RETURNED AND EVALUATED BY THE PRODUCT ANALYSIS LAB (PAL). THE EVENT HISTORY LOG REVIEW SHOWED NO KEYSTROKES, PROGRAMMING, OR USE RELATED EVENTS THAT INDICATED AND/OR CONTRIBUTED TO THE RITE (RETURNED INSTRUMENT TEST EVALUATION) FAILURE. THE EXTERNAL/INTERNAL INSPECTION WAS PERFORMED AND THE DEVICE PASSED, ALONG WITH ALL OF THE ELECTRICAL TESTING. DURING THE EVALUATION, THE FLUID WAS TRANSFERRED OUTSIDE OF THE RITE SPECIFIED LIMITS; THE VOLUMETRIC ACCURACY TEST THAT FOLLOWED SHOWED THAT THE DEVICE HAD FAILED. THE RESULTS OF THE EVALUATION REVEALED THE CAUSE OF THE FAILURE TO BE DETERIORATED PISTON FOAM. THE PISTON FOAM WAS TO BE SCRAPPED AND THE DEVICE WAS SENT FOR SERVICING. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.
(B)(4). THE DEVICE HAS BEEN RECEIVED AND THE EVALUATION IS IN PROGRESS. UPON COMPLETION OF THE DEVICE EVALUATION, OR IF ANY ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
DURING EVALUATION OF A RETURNED HOMECHOICE DEVICE, THE DEVICE FAILED FLUID VOLUME ACCURACY TESTING. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 379174 | HOMECHOICE | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - LARGO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |