PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 3004209178-2014-86389
- Event Type
- Malfunction
- Date Received
- June 30, 2014
- Date of Event
- May 10, 2014
- Report Date
- June 16, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX
- Reporter Occupation
- PATIENT
Narratives
FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP WAS RECEIVED WITH SCRATCHED SCREEN, CRACKED RESERVOIR TUBE, BROKEN RESERVOIR TUBE LIP, MISSING RESERVOIR TUBE LIP O-RING, CRACKED BATTERY TUBE THREADS, MISSING END CAP STICKER, AND LOOSE/PROTRUDED DRIVE SUPPORT DISK.
THE CUSTOMER REPORTED HAVING ISSUES WITH THE INSULIN PUMP. THE BLOOD GLUCOSE READING WAS 212MG/DL. THE CALLER STATED THAT THE DEVICE SEEMS TO BE WORKING, BUT HE WAS NOR RECEIVING ANY INSULIN. HE STATED THAT A PIECE OF PLASTIC AND THE BLACK RUBBER RING IN THE RESERVOIR COMPARTMENT CAME FROM THE INSULIN PUMP; IT HAPPENED WHEN HE WAS ATTEMPTING TO LOCK THE RESERVOIR INTO PLACE. ADVISED THE CALLER TO DISCONTINUE THE USE OF THE DEVICE AND REVERT TO BACK UP PLAN. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 379395 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-523NAL |