COULTER LH 780 HEMATOLOGY ANALYZER
Report
- Report Number
- 1061932-2014-01415
- Event Type
- Malfunction
- Date Received
- June 30, 2014
- Date of Event
- June 1, 2014
- Report Date
- June 1, 2014
- Manufacturer
- BECKMAN COULTER
- Product Code
- GKZ
- PMA / PMN Number
- K061616
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
A BECKMAN COULTER (BEC) FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO EVALUATE THE INSTRUMENT. THE FSE CONFIRMED THE LEAK AND ASPIRATION ERROR MESSAGES; THE ISSUES WERE RESOLVED BY REFITTING THE NEEDLE VENT TUBING AND REPLACING THE TUBING BETWEEN THE FIRST BLOOD DETECTOR AND NEEDLE ASSEMBLY. (B)(4).
THE CUSTOMER REPORTED A LEAK AND ASPIRATION ERROR MESSAGES FROM A COULTER LH 780 HEMATOLOGY ANALYZER. THE LEAK WAS NOT CONTAINED TO THE INSTRUMENT. THE OPERATOR WAS WEARING GLOVES AND A LAB COAT WHEN THE LEAK WAS FOUND. THERE WERE NO REPORTS OF DIRECT CONTACT WITH THE LEAK OR OF INJURY AS A RESULT OF THE LEAK. THE CUSTOMER REPLACED THE NEEDLE ASSEMBLY AND TUBING, BUT THE LEAK WAS NOT RESOLVED. THERE WERE NO ERRONEOUS RESULTS GENERATED IN CONNECTION WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 379165 | COULTER LH 780 HEMATOLOGY ANALYZER | COUNTER, DIFFERENTIAL CELL | GKZ | BECKMAN COULTER | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |