FDA Adverse Event
Malfunction
Summary report: N
G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM
MDR report key: 3903508
·
Received June 30, 2014
Report
- Report Number
- 3004753838-2014-24012
- Event Type
- Malfunction
- Date Received
- June 30, 2014
- Date of Event
- June 3, 2014
- Report Date
- June 3, 2014
- Manufacturer
- DEXCOM, INC.
- Product Code
- MDS
- PMA / PMN Number
- P120005
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PATIENT
Narratives
Description of Event or Problem · 1
PATIENT CONTACTED DEXCOM TECHNICAL SUPPORT ON (B)(6) 2014 TO REPORT CGM INACCURACIES COMPARED TO BLOOD GLUCOSE METER ON (B)(6) 2014. DEXCOM REVIEWED RECEIVER DATA ON (B)(6) 2014. DATA CONFIRMED SOME READINGS WERE OUTSIDE OF THE 20 PERCENT/20 POINT DIFFERENCE RANGEAT THE TIME OF THE CALL TO DEXCOM TECHNICAL SUPPORT, THE PATIENT DID NOT REPORT ANY MEDICAL INTERVENTION OR INJURIES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 379108 | G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM | MDS | MDS | DEXCOM, INC. | 9500-27 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 29 YR | Other |