ULTRACISION HARMONIC ACE
Report
- Report Number
- 3005075853-2014-04510
- Event Type
- Malfunction
- Date Received
- June 30, 2014
- Date of Event
- June 23, 2014
- Report Date
- June 24, 2014
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- LFL
- PMA / PMN Number
- K042777
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- OTHER
Narratives
(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME. NO DEVICE RECEIVED FOR ANALYSIS AT TIME OF SUBMISSION OF 3500A. WHEN ADDITIONAL INFORMATION IS RECEIVED AND/OR THE DEVICE ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL MEDWATCH WILL BE SENT.
(B)(4). ADDITIONAL INFORMATION: THE DEVICE WAS RETURNED WITH THE BLADE TIP BROKE OFF AND RETURNED. DURING FUNCTIONAL TESTING ON GEN11 AN ALERT SCREEN WAS DISPLAYED. THE DEVICE WILL STOP ACTIVATING WHEN THE BLADE BECOMES DAMAGED. THE DEVICE WAS DISASSEMBLED AND BLADE WAS BROKEN INSIDE THE TUBE PROXIMAL TO THE TISSUE PAD. ONCE MINOR BLADE DAMAGE HAS OCCURRED, SUBSEQUENT ACTIVATIONS MAY INCREASE DAMAGE SEVERITY AND RESULT IN ALERT SCREENS, SUCH AS ¿TIGHTEN ASSEMBLY¿, ¿BLADE ERROR DETECTED¿ OR "RELAXED PRESSURE ON BLADE" FOLLOWED BY A ¿REPLACE INSTRUMENT¿ SCREEN LATER IN THE PROCEDURE. THE ANALYSIS CONCLUDED THAT THE BLADE BROKE DUE TO CONTACT WITH CLAMP ARM PIN. NO CONCLUSION COULD BE REACHED AS TO WHAT CAUSED THE BLADE TO MAKE CONTACT WITH THE CLAMP ARM PIN.
IT WAS REPORTED THAT DURING A LAPAROSCOPIC SLEEVE PROCEDURE (JAW) ONE TIP IS BROKEN. NO FURTHER INFORMATION PROVIDED. NO PATIENT CONSEQUENCES REPORTED. ONE DEVICE WILL BE RETURNING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 379650 | ULTRACISION HARMONIC ACE | INSTRUMENT, ULTRASONIC SURGICAL | LFL | ETHICON ENDO-SURGERY, LLC. | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | GENERATOR AND HANDPIECE |