FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3903426 · Received June 30, 2014

Report

Report Number
3004209178-2014-86188
Event Type
Malfunction
Date Received
June 30, 2014
Date of Event
June 13, 2014
Report Date
June 13, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

THE INSULIN PUMP ALARMED DURING THE BASIC OCCLUSION DUE TO PROTRUDED/LOOSE DRIVE SUPPORT DISK. THE INSULIN PUMP RECEIVED WITH MINOR SCRATCHES ON DISPLAY WINDOW, CRACKED CASE ON DISPLAY WINDOW CORNER, CRACKED BATTERY TUBE THREADS, CRACKED RESERVOIR TUBE LIP AND CRACKED BELT CLIP SLOT.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP ALARMED DURING THE PRIME PROCESS. INSULIN SQUIRTED OUT DURING THE MANUAL PRIME PROCESS. THE CURRENT BLOOD GLUCOSE READING IS 173 MG/DL. THE DRIVE SUPPORT CAP IS PROTRUDED. THE INSULIN PUMP WILL BE REPLACED. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
379436 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAS

Patients

Seq Age Sex Outcome Treatment
1 25 YR