FDA Adverse Event
Malfunction
Summary report: N
V60 VENTILATOR
MDR report key: 3903327
·
Received June 30, 2014
Report
- Report Number
- 2031642-2014-00583
- Event Type
- Malfunction
- Date Received
- June 30, 2014
- Report Date
- June 4, 2014
- Manufacturer
- RESPIRONICS CALIFORNIA, INC.
- Product Code
- MNT
- PMA / PMN Number
- K082660
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
Description of Event or Problem · 1
FACTORY ANALYSIS OF A RETURNED POWER MANAGEMENT PCB BOARD REVEALED A DIODE FAILURE THAT MAY RESULT IN THE UNIT SHUTTING DOWN DURING USE IN NORMAL VENTILATION MODE OPERATION. UPON LOSS OF POWER, A POWER FAIL ALARM WILL ACTIVATE AND IF IN USE ALTERNATIVE VENTILATION SHOULD BE PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 379524 | V60 VENTILATOR | VENTILATOR, CONTINUOUS | MNT | RESPIRONICS CALIFORNIA, INC. | V60 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |