FLOGARD
Report
- Report Number
- 1416980-2014-20899
- Event Type
- Malfunction
- Date Received
- June 30, 2014
- Date of Event
- June 5, 2014
- Report Date
- June 5, 2014
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MX
- Reporter Occupation
- OTHER
Narratives
(B)(4). EVALUATION SUMMARY: THE DEVICE WAS SERVICED ON-SITE BY A FIELD SERVICE TECHNICIAN. VISUAL INSPECTION AND FUNCTIONAL TESTING WERE PERFORMED. FUNCTIONAL FORCE SENSING RESISTOR TESTING REVEALED THAT THE DEVICE HAD A F-38 ALARM. THE CAUSE OF THE CONDITION WAS DETERMINED TO BE AN INOPERATIVE FORCE SENSING RESISTOR. TO CORRECT THE CONDITION, THE FORCE SENSING RESISTOR WAS REPLACED. THE DEVICE WAS SERVICED AND RESTORED TO A GOOD WORKING CONDITION. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
IT WAS REPORTED THAT A FLOGARD INFUSION PUMP HAD A F-38 ALARM. IT WAS NOT SPECIFIED WHEN IN THE PROCESS THIS OCCURRED. THERE WAS NO REPORT OF PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 380174 | FLOGARD | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |