FDA Adverse Event Malfunction Summary report: N

FLOGARD

MDR report key: 3903000 · Received June 30, 2014

Report

Report Number
1416980-2014-20899
Event Type
Malfunction
Date Received
June 30, 2014
Date of Event
June 5, 2014
Report Date
June 5, 2014
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MX
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE DEVICE WAS SERVICED ON-SITE BY A FIELD SERVICE TECHNICIAN. VISUAL INSPECTION AND FUNCTIONAL TESTING WERE PERFORMED. FUNCTIONAL FORCE SENSING RESISTOR TESTING REVEALED THAT THE DEVICE HAD A F-38 ALARM. THE CAUSE OF THE CONDITION WAS DETERMINED TO BE AN INOPERATIVE FORCE SENSING RESISTOR. TO CORRECT THE CONDITION, THE FORCE SENSING RESISTOR WAS REPLACED. THE DEVICE WAS SERVICED AND RESTORED TO A GOOD WORKING CONDITION. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A FLOGARD INFUSION PUMP HAD A F-38 ALARM. IT WAS NOT SPECIFIED WHEN IN THE PROCESS THIS OCCURRED. THERE WAS NO REPORT OF PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
380174 FLOGARD PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1