TX1 TISSUE REMOVAL SYSTEM - MICROTIP
Report
- Report Number
- 2085033-2014-00548
- Event Type
- Malfunction
- Date Received
- February 24, 2014
- Date of Event
- January 28, 2014
- Report Date
- February 14, 2014
- Manufacturer
- TENEX HEALTH
- Product Code
- LFL
- PMA / PMN Number
- K123640
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
ON (B)(4) 2014 UNIT WAS RECEIVED FROM THE FIELD FOR EVALUATION. (B)(4). NO VISUAL DEFECT NOTED. PRIMED SUCCESSFULLY, CUT FOR 18 SECONDS AND THEN GAVE "CHECK HANDPIECE" ERROR MESSAGE (TEST002). UPON INTERNAL INSPECTION A BROKEN JUMPER WAS DETECTED. HANDPIECE WAS DISCARDED. ON (B)(6) 2014 THE DHR FOR MICROTIP LOT 33813-05 WAS REVIEWED AND SHOWED THAT THE UNITS MET MANUFACTURING SPECIFICATIONS WHEN RELEASED AND THERE WERE NO ISSUES RELATED TO THE COMPLAINT. THERE WERE (B)(4) UNITS RELEASED FROM MICROTIP LOT 33813-05. A REVIEW OF LOT HISTORY SHOWED THAT THERE WAS ONLY ONE (1) REPORTED MICROTIP UNIT FAILURE FROM THIS LOT. LOT 33813-05 HAS A FAILURE RATE OF (B)(4). NO SIGNIFICANT TREND WAS OBSERVED. A REVIEW OF CONSOLE HISTORY ((B)(4)) WAS PERFORMED DURING THE MONTHS OF 08/29/2013 THRU 01/29/2014 AND SHOWED THAT THERE WAS ONLY ONE REPORTED COMPLAINT ORIGINATING FROM THE CONSOLE. NO SIGNIFICANT TREND WAS OBSERVED. THE COMPLAINTS DATABASE WAS REVIEWED DURING THE MONTHS OF 08/29/2013 THROUGH 01/29/2014 AND SHOWED THAT THIS WAS THE FIRST AND ONLY COMPLAINT REPORTED FROM THIS FACILITY. NO SIGNIFICANT TREND WAS OBSERVED.
WE USED THE TX MICROTIP FOR ABOUT 30 SECONDS OF POWER DURING THE CASE. IT THEN STOPPED WORKING. DISPLAY RED "CHECK HANDPIECE." TRIED ALL TROUBLESHOOTING. NOTHING WORKED. HAD TO SWAP OUT FOR A NEW HANDPIECE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 113556 | TX1 TISSUE REMOVAL SYSTEM - MICROTIP | ULTRASONIC SURGICAL ASPIRATOR | LFL | TENEX HEALTH | 33813-05 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |