FDA Adverse Event Malfunction Summary report: N

HOMECHOICE PRO

MDR report key: 3902886 · Received June 30, 2014

Report

Report Number
1416980-2014-20879
Event Type
Malfunction
Date Received
June 30, 2014
Date of Event
January 18, 2014
Report Date
June 5, 2014
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED TO BAXTER AND THE EVALUATION IS COMPLETE. THIS COMPLAINT IS AN ANCILLARY SERVICE EVENT. ALL RETURNED HOMECHOICE DEVICES RECEIVE A RETURNED INSTRUMENT TESTING EVALUATION (RITE). THIS EVALUATION INCLUDES FUNCTIONAL AND ELECTRICAL TESTING ON THE DEVICE. THE DEVICE ALSO RECEIVED AN INTERNAL/EXTERNAL INSPECTION. THE HIGH DRAIN 105 EVENT WAS VERIFIED THROUGH AN EVENT HISTORY LOG REVIEW. THE CAUSE WAS UNDETERMINED AS NOT ENOUGH EVIDENCE WAS AVAILABLE TO MAKE A DETERMINATION. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED HOMECHOICE DEVICE, A HIGH DRAIN ERROR 105 (NIGHT DRAIN CYCLE 5) ALARM WAS IDENTIFIED IN THE LOG (INCLUDES ANY STANDARD MODE THERAPY WHERE THE PATIENT DRAIN VOLUME EXCEEDS 200% OF THE MAXIMUM PRESCRIBED CYCLE FILL VOLUME). THE ALARM OCCURRED ON (B)(6) 2014 05:54:24. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
378973 HOMECHOICE PRO SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1 62 YR