FDA Adverse Event Malfunction Summary report: N

2008H MACHINE, SHORT CAB, STD, WITH H.P.

MDR report key: 3902811 · Received February 14, 2014

Report

Report Number
2937457-2014-00234
Event Type
Malfunction
Date Received
February 14, 2014
Date of Event
January 24, 2014
Report Date
February 4, 2014
Manufacturer
FRESENIUS MEDICAL CARE NORTH AMERICA
Product Code
KDI
PMA / PMN Number
K994267
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION FINDINGS TO DATE INDICATE THAT THE REPORTED MALFUNCTION OCCURRED DURING RECIRCULATION AND PRIME (MACHINE SET-UP), AND NOT DURING DIALYSIS MODE. THE USER VISUALLY OBSERVED THE SALINE BAG REFILLING WITH DIALYSATE DURING CIRCULATION. THERE HAVE BEEN NO ADVERSE EVENTS ASSOCIATED WITH THE REPORTED ISSUE. THE REPORT IS BEING INVESTIGATED BY THE MANUFACTURER VIA A CAPA. THE INVESTIGATION IS PENDING A SUPPLEMENTAL MDR WILL BE FILED AT THE COMPLETION OF THE INVESTIGATION.

Description of Event or Problem · 1

A USER FACILITY REPORTED A SALINE BAG BACK FILLED DURING RECIRCULATION MODE. A PATIENT WAS NOT CONNECTED TO THE MACHINE AT THE TIME OF THE INCIDENT. THERE WERE NO PARTS REPLACED ON THE DEVICE BY THE TECHNICIAN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
98461 2008H MACHINE, SHORT CAB, STD, WITH H.P. KDI FRESENIUS MEDICAL CARE NORTH AMERICA H

Patients

Seq Age Sex Outcome Treatment
1 UNK SALINE