FDA Adverse Event
Malfunction
Summary report: N
9900
MDR report key: 3902682
·
Received May 9, 2014
Report
- Report Number
- 1720753-2014-04057
- Event Type
- Malfunction
- Date Received
- May 9, 2014
- Date of Event
- April 30, 2014
- Report Date
- May 9, 2014
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE SOFTWARE WAS INSTALLED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE SYSTEM WOULD NOT BOOT UP DUE TO CORRUPT SOFTWARE. THIS RESULTED IN A TOTAL LOSS OF IMAGING FUNCTIONALITY. THIS COULD CAUSE THE DELAY OR CANCELLATION OF A PROCEDURE. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 282289 | 9900 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |