FDA Adverse Event Malfunction Summary report: N

HS III PROXIMAL SEAL

MDR report key: 3902613 · Received May 9, 2014

Report

Report Number
2242352-2014-00533
Event Type
Malfunction
Date Received
May 9, 2014
Date of Event
April 8, 2014
Report Date
April 15, 2014
Manufacturer
MAQUET CARDIOVASCULAR, LLC
Product Code
DXC
PMA / PMN Number
K080169
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RETURNED TO THE FACTORY FOR EVAL. IT SHOWED NO SIGNS OF CLINICAL USAGE OR EVIDENCE OF BLOOD. THE DELIVERY DEVICE WAS RETURNED OUTSIDE THE LOADING DEVICE. THE TENSION SPRING ASSEMBLY, SEAL AND ANCHOR TAB WERE INSIDE THE BODY OF THE LOADING DEVICE. THE BLUE SLIDE LOCK WAS LOCKED AND THE WHITE PLUNGER WAS NOT DEPRESSED ON THE DELIVERY DEVICE. BASED UPON THE RECEIVED CONDITION OF THE DEVICE, THE REPORTED COMPLAINT WAS CONFIRMED FOR FAILURE TO LOAD. A LOT HISTORY RECORD REVIEW WAS COMPLETED FOR THE REPORTED PRODUCT LOT NUMBER. THERE WAS NO NON-CONFORMANCE RECORDED IN THE LOT HISTORY. (B)(4).

Description of Event or Problem · 1

THE HOSPITAL REPORTED THAT DURING A CORONARY ARTERY BYPASS PROCEDURE, WHILE SQUEEZING THE GREEN BUTTONS ON THE LOADER OF THE HEARTSTRING III THE SEAL DID NOT SET PROPERLY. THE SEAL REMAINED INSIDE THE LOADER WHEN THE DELIVERY DEVICE WAS PULLED. A REPLACEMENT DEVICE WAS USED TO COMPLETE THE PROCEDURE. THE HOSPITAL DID NOT REPORT ANY PT EFFECTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
282351 HS III PROXIMAL SEAL ENDOSCOPIC VESSEL HARVESTING DXC MAQUET CARDIOVASCULAR, LLC HS-3045 25087631

Patients

Seq Age Sex Outcome Treatment
1