ENDOTRACHEAL TUBES - AIRWAY MANAGEMENT
Report
- Report Number
- 9611710-2014-00100
- Event Type
- Malfunction
- Date Received
- May 7, 2014
- Date of Event
- April 7, 2014
- Report Date
- April 7, 2014
- Manufacturer
- UNOMEDICAL SDN BHD
- Product Code
- BTR
- PMA / PMN Number
- K951696
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- NOT APPLICABLE
Narratives
BASED ON THE AVAILABLE INFO, THIS EVENT IS DEEMED A REPORTABLE MALFUNCTION. AN EVALUATION WAS PERFORMED ON 05/05/2014. RECEIVED (B)(4) UNUSED I.D 9.5MM, CATALOG#: 61110095 ENDOTRACHEAL TUBES WITH FROSTY BALLOONS, FITTED WITH PP CONNECTORS OF CORRESPONDING SIZES AND TUBES MARKED AS PER JSS SPECIFICATION. PRODUCTS FITTED WITH PVC BULLET VALVES WITH PILOT BALLOONS PRINTED WITH SIZES IDENTIFICATION 9.5 MM AND WORK ORDER #: (B)(4). THE PRODUCTS WERE REC'D IN SEALED JSS PREPRINTED BLISTER PACK AND AFFIXED WITH LABEL. THE SAMPLES WERE INSPECTED AND THE OBSERVATIONS ARE AS FOLLOWS: FOUR UNITS CONFIRMED FOR FOREIGN SUBSTANCE IN PACKAGE, WHEREAS ONE UNIT SHOWED NO SIGNS OF FOREIGN SUBSTANCE IN PACKAGE; TWO UNITS CONFIRMED FOR STAIN ON PRODUCT; TWO UNITS CONFIRMED FOR PRODUCT BROKEN AND NINETEEN UNITS CONFIRMED FOR DEFORMED. REVIEW OF THE ASSEMBLY RECORD DID NOT REVEAL ANY SIGNS OF THE ASSOCIATED DEFECTS DETECTED DURING THE (B)(4) LIGHT TEST. BASED ON THE ANALYSIS OF THE RETURNED SAMPLES PROVIDED IT WAS FOUND THAT (B)(4) OUT OF (B)(4) UNITS RETURNED WERE NOT MEETING OUR INSPECTION CRITERIA. HOWEVER, THIS COMPLAINT ISSUE HAS BEEN INVESTIGATED TO ADDRESS THE CONFIRMED COMPLAINT AND DOCUMENTED IN THE CAPA SYSTEM. NO ADD'L PATIENT/EVENT DETAILS HAVE BEEN PROVIDED TO DATE. SHOULD ADD'L INFO BECOME AVAILABLE A F/U REPORT WILL BE SUBMITTED.
IT WAS REPORTED THAT DURING INCOMING INSPECTION "FOUND CONTAMINATION ON PACKAGE".
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 276857 | ENDOTRACHEAL TUBES - AIRWAY MANAGEMENT | TUBE, TRACHEAL (W/WO CONNECTOR) | BTR | UNOMEDICAL SDN BHD | 61110095 | 616034R001 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |