FDA Adverse Event Malfunction Summary report: N

EXPRESS 4

MDR report key: 3902445 · Received May 6, 2014

Report

Report Number
2023446-2014-00052
Event Type
Malfunction
Date Received
May 6, 2014
Date of Event
April 30, 2014
Report Date
April 30, 2014
Manufacturer
IRIS INTERNATIONAL
Product Code
JQC
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OK, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

STATSPIN CENTRIFUGE HAD 120VAC AT 24V PS IN AND 24VDS AT 24VPS OUT, BUT THERE WAS NO LIGHTS ON THE PANEL. IT WAS DETERMINED THAT POWER HARNESS AT CONTROL BOARD WAS BURNED. NO INJURIES REPORTED. NO VISIBLE FLAMES, SMOKE OR BURNING SMELL REPORTED, NOR WAS FIRE DEPT CALLED, BUT CHARRED MATERIAL PRESENT.

Description of Event or Problem · 1

CUSTOMER REPORTED TO 120VAC AT 24VPS IN AND 24VDC AT 24VPS OUT, BUT THERE WERE NO LIGHTS ON THE PANEL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
272126 EXPRESS 4 STATSPIN SENTRIFUGE JQC IRIS INTERNATIONAL

Patients

Seq Age Sex Outcome Treatment
1