FDA Adverse Event
Malfunction
Summary report: N
2800
MDR report key: 3902421
·
Received May 6, 2014
Report
- Report Number
- 1720753-2014-03987
- Event Type
- Malfunction
- Date Received
- May 6, 2014
- Date of Event
- April 18, 2014
- Report Date
- May 6, 2014
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AR, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION. THE KV CONTROL PCB, GENERIC INTERFACE AND AEC PCB WERE EVALUATED AND REPLACED. THE AEC CALIBRATION FILED WERE ALSO RELOADED AS PART OF THE SERVICE EVENT. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THERE WAS AN "GENERATOR ERROR" MESSAGE DISPLAYED. THIS ERROR IS LIKELY TO RESULT IN AN IMMEDIATE LOSS OF SYSTEM FUNCTIONALITY AND AN UN-COMMANDED SHUT DOWN. THERE IS NO REPORT OF A PT INJURY OR DEATH ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 272831 | 2800 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 2800 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |