FDA Adverse Event Malfunction Summary report: N

6600

MDR report key: 3902420 · Received May 6, 2014

Report

Report Number
1720753-2014-03986
Event Type
Malfunction
Date Received
May 6, 2014
Date of Event
April 25, 2014
Report Date
May 6, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION. THE WORKSTATION CABLES AND CONNECTORS WERE EVALUATED AND RESEATED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM FAILED TO ALLOW FLUOROSCOPIC EXPOSURES AND EXHIBITED A NON-RECOVERABLE LOSS OF FUNCTIONALITY. NO PT SERIOUS INJURY OR DEATH WAS REPORTED RELATED TO THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
272439 6600 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 6600