VICRYL PLUS ANTIBACTERIAL (POLYGLACTIN 910) SUTURE
Report
- Report Number
- 2210968-2014-08421
- Event Type
- Malfunction
- Date Received
- June 30, 2014
- Date of Event
- May 26, 2014
- Report Date
- June 10, 2014
- Manufacturer
- ETHICON INC.
- Product Code
- GAM
- PMA / PMN Number
- K032420
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
(B)(4). CONCLUSION: TO DATE, THE DEVICE HAS NOT BEEN RETURNED. IF THE PRODUCT IS RETURNED FOR EVALUATION, ANY FURTHER INFO DERIVED FROM THE EVALUATION WILL BE SUBMITTED IN A SUPPLEMENTAL 3500A FORM.
CONCLUSION: A NEEDLE THAT BROKE IN THE BODY TOWARDS THE ATTACHMENT END WAS SUBMITTED FOR THIS EVALUATION. A MICROSCOPIC INSPECTION REVEALED INDENTS AND SCUFF MARKS AROUND THE BREAK AREA PRODUCED DURING HANDLING BY THE SURGICAL NEEDLE HOLDERS OR SOME OTHER GRIPPING DEVICE. THE NEEDLE FRACTURED DUE TO TENSILE OVERLOAD GENERATED DURING SEVERE MECHANICAL DEFORMATION, WITH SIGNS OF DUCTILITY.
A REVIEW OF THE BATCH MANUFACTURING RECORDS WAS CONDUCTED AND THE BATCH MET ALL FINISHED GOODS RELEASE CRITERIA.
IT WAS REPORTED THAT A PATIENT UNDERWENT A SURGICAL PROCEDURE ON (B)(6) 2014 AND SUTURE WAS USED. DURING THE PROCEDURE, THE NEEDLE BROKE DURING SUTURING. THE NEEDLE WAS RETRIEVED. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WERE NO ADVERSE PATIENT CONSEQUENCES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 380695 | VICRYL PLUS ANTIBACTERIAL (POLYGLACTIN 910) SUTURE | SUTURE, ABSORBABLE, SYNTHETIC | GAM | ETHICON INC. | UNK | HC2319 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |