FDA Adverse Event Malfunction Summary report: N

TX1 TISSUE REMOVAL SYSTEM - MICROTIP

MDR report key: 3902346 · Received February 24, 2014

Report

Report Number
2085033-2014-00074
Event Type
Malfunction
Date Received
February 24, 2014
Report Date
February 16, 2014
Manufacturer
AMERICAN OPTISURGICAL, INC.
Product Code
LFL
PMA / PMN Number
K123640
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THERE IS NO SERVICE RECORD, AND WITHOUT EVALUATING THE UNIT, THE CAUSE FOR MALFUNCTION CANNOT BE DETERMINED. ON (B)(4) 2012: AS OF THIS DATE, THERE ARE 4 DOCUMENTED DEVICE COMPLAINTS COVERING 6 UNITS. THIS LOT HAS (B)(4) RELEASED UNITS. THE COMPLAINT RATE IS AT (B)(4). THE LOT WILL CONTINUE TO BE MONITORED. NO CAPA WILL BE ISSUED SINCE THE COMPLAINT RATE IS BELOW (B)(4) AND THE INCIDENCES CAN BE CONSIDERED OPERATOR INEXPERIENCE. ON (B)(4) 2014: OPENED CAPA (B)(4) TO REVIEW 2012 COMPLAINTS. UPDATED WITH THE FOLLOWING DETAILS: THE UNIT HAD MADE PT CONTACT AND WAS NOT RECEIVED FOR EVALUATION. IT IS NOT POSSIBLE TO DETERMINE AN ASSIGNABLE CAUSE FOR THIS ISSUE. THERE IS INSUFFICIENT INFO FOR A COMPETE INVESTIGATION. COMPLETED AND ATTACHED A CURRENT MDR DECISION FLOWCHART. THE DHR FOR LOT # 12912-04 WAS REVIEWED. THE UNIT MET MFG SPECIFICATIONS WHEN RELEASED AND THERE WERE NO ISSUES RELATED TO THE COMPLAINT. THIS LOT HAS A (B)(4) COMPLAINT RATE. THERE WERE (B)(4) UNITS RELEASED TO DISTRIBUTION. A REVIEW OF LOT HISTORY SHOWED THAT THERE WERE A TOTAL OF 2 UNITS THAT STOPPED WORKING. A REVIEW OF THE CONSOLE HISTORY WAS NOT PERFORMED BECAUSE THE CONSOLE SERIAL NUMBER WAS NOT REPORTED WITH THE COMPLAINT.

Description of Event or Problem · 1

HANDPIECE WAS ASPIRATING THEN STATED "CHECK HANDPIECE".

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
113560 TX1 TISSUE REMOVAL SYSTEM - MICROTIP ULTRASONIC SURGICAL ASPIRATOR LFL AMERICAN OPTISURGICAL, INC. 12912-04

Patients

Seq Age Sex Outcome Treatment
1