FDA Adverse Event Malfunction Summary report: N

ES6 SINGLE TRIGGER ROTARY

MDR report key: 3902334 · Received June 30, 2014

Report

Report Number
0001811755-2014-02322
Event Type
Malfunction
Date Received
June 30, 2014
Date of Event
June 3, 2014
Report Date
June 3, 2014
Manufacturer
STRYKER INSTRUMENTS-KALAMAZOO
Product Code
HWE
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE REPORTED EVENT WAS DUPLICATED. UPON DISASSEMBLY THE TECHNICIAN NOTED A FAILURE OF THE POTTED TRIGGER PCB ASSEMBLY.

Description of Event or Problem · 1

THE ES6 SINGLE TRIGGER ROTARY WAS RETURNED TO STRYKER INSTRUMENTS FOR SERVICE. DURING FUNCTIONAL TESTING BY A SERVICE TECHNICIAN, IT WAS FOUND THAT THE DEVICE CAUSED A BIAS CURRENT MESSAGE TO BE DISPLAYED ON THE CONSOLE, SIGNALING A CONDITION OCCURRED IN WHICH THE DEVICE HAS THE POTENTIAL TO RUN WITHOUT USER ACTIVATION. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING TESTING AT THE MANUFACTURER FACILITY, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.