FDA Adverse Event Malfunction Summary report: N

SYMBIQ SCHANNEL 3.01

MDR report key: 3902110 · Received May 2, 2014

Report

Report Number
9615050-2014-03163
Event Type
Malfunction
Date Received
May 2, 2014
Date of Event
January 1, 2014
Report Date
April 4, 2014
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K110901
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

TESTING AND INVESTIGATION FOUND THE DEVICE DID NOT ALARM WITH A S321 (MOTOR ERROR-PMC, LEFT) MALFUNCTION ALARM CODE. DURING REVIEW OF THE DEVICE HISTORY S321 MALFUNCTION ALARMS WERE NOTED. THE PROBABLE CAUSE OF THE S321 MALFUNCTION ALARM CODE IS A BROKEN ENCODER ON THE MOTOR. THIS REPORT REPRESENTS ALL THE INFO KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

THE CUSTOMER CONTACT REPORTED THE DEVICE ALARMED WITH A S321 (MOTOR ERROR - PMC, LEFT) SERVICE ALARM CODE. THE DEVICE WAS RETURNED TO THE BIOMEDICAL DEPARTMENT FOR AN UNSPECIFIED REASON. NO INFO WAS PROVIDED; THEREFORE, SPECIFIC PT INFO, PUMP PROGRAMMING, OR EVENT DETAILS WERE NOT AVAILABLE. THERE WERE NO REPORTED OF ANY ADVERSE PT EVENTS AND NO REPORTED DELAYS OF CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. DURING TESTING AT THE FACILITY, THE DEVICE ALARMED WITH A S321 SERVICE ALARM CODE. NO ADDITIONAL INFO WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
265319 SYMBIQ SCHANNEL 3.01 80FRN FRN HOSPIRA COSTA RICA LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1 UNK