FDA Adverse Event
Malfunction
Summary report: N
SYMBIQ SINGLE CHANNEL
MDR report key: 3902089
·
Received May 2, 2014
Report
- Report Number
- 9615050-2014-03169
- Event Type
- Malfunction
- Date Received
- May 2, 2014
- Date of Event
- April 1, 2014
- Report Date
- April 2, 2014
- Manufacturer
- HOSPIRA COSTA RICA LTD.
- Product Code
- FRN
- PMA / PMN Number
- K110901
- Removal / Correction Number
- Z-0069-2013
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
TESTING AND INVESTIGATION FOUND THE TOUCHSCREEN DID NOT RESPOND. THE PROBABLE CAUSE WAS DUE TO CONTAMINATION ON THE BOTTOM OF THE TOUCHSCREEN DUE TO FLUID INGRESS. THIS DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFO KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.
Description of Event or Problem · 1
THE CUSTOMER CONTACT REPORTED THE DEVICE TOUCHSCREEN SEEMS BROKEN, DOES NOT RESPOND. THERE WERE NO REPORTS OF ANY ADVERSE PT EVENTS AND NO REPORTED DELAYS OF CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. DURING TESTING AT THE USER FACILITY, THE DEVICE TOUCHSCREEN DOES NOT RESPOND WHEN PRESSED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 265614 | SYMBIQ SINGLE CHANNEL | 80FRN | FRN | HOSPIRA COSTA RICA LTD. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |