FDA Adverse Event Malfunction Summary report: N

GEMSTR 7 THPY ALT ST

MDR report key: 3902032 · Received May 7, 2014

Report

Report Number
9615050-2014-03291
Event Type
Malfunction
Date Received
May 7, 2014
Date of Event
March 17, 2014
Report Date
March 17, 2014
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K060806
Removal / Correction Number
Z-1159-2013, Z-1169-2013
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DURING TESTING AT POWER UP, THE DEVICE ALARMED FOR 11/004 (LESS THAN 2.0 VOLTS ON LITHIUM BATTERY) SERVICE CODE ALARM. DURING FURTHER TESTING, LEAKAGE WAS NOTED FROM THE DISPOSABLE BATTERIES IN THE BATTERY COMPARTMENT OF THE DEVICE. THE CUSTOMER CONTACT'S REPORT OF A BOTTOM CAP WAS FOUND TO BE A BROKEN BOTTOM END CAP. THE DEVICE HAS BEEN IDENTIFIED AS PART OF TWO PRODUCT RECALLS. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

THE DEVICE WAS RETURNED TO THE SERVICE CENTER WITH A REPORT OF "BOTTOM CAP." THIS IS NOT A REPORTABLE MALFUNCTION. HOWEVER, DURING VERIFICATION TESTING AT THE SERVICE CENTER, THE DEVICE ALARMED WITH A 11/004 (LESS THAN 2.0 VOLTS ON LITHIUM BATTERY) SERVICE ALARM CODE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
276766 GEMSTR 7 THPY ALT ST 80FRN FRN HOSPIRA COSTA RICA LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1 NA