FDA Adverse Event Malfunction Summary report: N

PLUM XLD 110V L.A.

MDR report key: 3902028 · Received May 7, 2014

Report

Report Number
9615050-2014-03287
Event Type
Malfunction
Date Received
May 7, 2014
Date of Event
January 1, 2014
Report Date
April 14, 2014
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K010924
Removal / Correction Number
FA311-04
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A FIELD SERVICE ENGINEER PERFORMED TESTING AND INVESTIGATION AT THE USER FACILITY. DURING TESTING, IT WAS NOTED THAT THE DEVICE DOOR ROLLER PIN WAS MISSING. THE DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFO KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

DURING TESTING AT THE USER FACILITY, IT WAS NOTED THAT THE DOOR ROLLER PIN WAS MISSING. THE DEVICE HAD BEEN RETURNED TO THE BIOMEDICAL DEPARTMENT WITH A REPORT THAT THE DEVICE DOES NOT RECOGNIZE THE CASSETTE. NO TRACKING INFO WAS PROVIDED; THEREFORE, SPECIFIC PT INFO, DEVICE PROGRAMMING, OR EVENT DETAILS WERE NOT AVAILABLE. THERE WERE NO REPORTS OF ANY ADVERSE PT EVENTS OR DELAYS IN CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. NO ADD'L INFO WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
276872 PLUM XLD 110V L.A. 80FRN FRN HOSPIRA COSTA RICA LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1 UNK