FDA Adverse Event Malfunction Summary report: N

HS III PROXIMAL SEAL

MDR report key: 3901737 · Received May 5, 2014

Report

Report Number
2242352-2014-00505
Event Type
Malfunction
Date Received
May 5, 2014
Date of Event
April 2, 2014
Report Date
April 8, 2014
Manufacturer
MAQUET CARDIOVASCULAR, LLC
Product Code
DXC
PMA / PMN Number
K080169
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RETURNED TO THE FACTORY FOR EVALUATION. IT SHOWED NO SIGNS OF CLINICAL USAGE OR EVIDENCE OF BLOOD. THE LOADING DEVICE WAS NOT RETURNED. THE TENSION SPRING ASSEMBLY AND ANCHOR TAB WERE INSIDE THE TUBE OF THE DELIVERY DEVICE. THE SEAL WAS EXTENDED OUTSIDE OF THE DELIVERY TUBE; IT REMAINED ANCHORED ON THE TENSION SPRING ASSEMBLY. THE GREEN SLIDE LOCK WAS LOCKED AND WHITE PLUNGER WAS NOT DEPRESSED ON THE DELIVERY DEVICE. BASED UPON THE EVALUATION RESULTS, THE REPORTED COMPLAINT WAS CONFIRMED FOR FAILURE TO LOAD. ADDITIONAL EVALUATION REQUIRED. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETED. A LOT HISTORY RECORD REVIEW COULD NOT BE PERFORMED AS THE PRODUCT LOT NUMBER IS UNKNOWN. (B)(4).

Description of Event or Problem · 1

THE HOSPITAL REPORTED THAT DURING PREPARATION FOR A CORONARY ARTERY BYPASS PROCEDURE, WHILE LOADING THE HEARTSTRING III THE SCRUB TECH OBSERVED A "CRIMP" IN THE PROXIMAL SEAL WHERE THE OUTER RING WAS UNRAVELING. A REPLACEMENT DEVICE WAS USED TO COMPLETE THE PROCEDURE. THE HOSPITAL DID NOT REPORT ANY PATIENT EFFECTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
268115 HS III PROXIMAL SEAL ENDOSCOPIC VESSEL HARVESTING DXC MAQUET CARDIOVASCULAR, LLC HS-3045

Patients

Seq Age Sex Outcome Treatment
1