FDA Adverse Event
Malfunction
Summary report: N
SELF RETAINING TORX DRIVER FOR 2.7 MM SURFIX NEWDEAL TIP
MDR report key: 3901640
·
Received May 6, 2014
Report
- Report Number
- 3004608878-2014-00054
- Event Type
- Malfunction
- Date Received
- May 6, 2014
- Date of Event
- April 14, 2014
- Report Date
- April 14, 2014
- Manufacturer
- INTEGRA LIFESCIENCES CORPORATION
- Product Code
- HRS
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE DEVICE INVOLVED IN THE REPORTED INCIDENT HAS BEEN RECEIVED FOR EVALUATION. AN INVESTIGATION HAS BEEN INITIATED BASED ON THE REPORTED INFORMATION.
Description of Event or Problem · 1
IT WAS REPORTED DURING A WRIST FUSION SURGERY, THE SURGEON INSERTED A 2.7 SURFIX SCREW ((B)(4)) INTO THE MOST DISTAL HOLE ON THE WRIST FUSION PLATE ((B)(4)) AFTER HE DRILLED WITH THE 2.0 DRILL BIT ((B)(4)). WHEN TIGHTENING THE SCREW DOWN (WITH THE SCREW DRIVE) THE TIP OF THE SCREW DRIVE ((B)(4)) BROKE OFF INSIDE THE SCREW. THE SMALL FRAGMENT OF SCREW DRIVER TIP WAS NOT ABLE TO BE REMOVED. IT WAS LEFT INSIDE OF THE HEAD OF THE 3.7 SCREW. THE SURGEON WAS UNABLE TO PUT A LOCK WASHER ON TOP OF THE SCREW TO LOCK IT. THERE WAS NO REPORTED INJURY OR ADVERSE CONSEQUENCE TO THE PATIENT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 271596 | SELF RETAINING TORX DRIVER FOR 2.7 MM SURFIX NEWDEAL TIP | WRIST FUSION INSTRUMENTS | HRS | INTEGRA LIFESCIENCES CORPORATION | JG0034 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 91 YR |