FDA Adverse Event Malfunction Summary report: N

TRILOGY 200

MDR report key: 3901568 · Received May 7, 2014

Report

Report Number
2518422-2014-00719
Event Type
Malfunction
Date Received
May 7, 2014
Date of Event
April 15, 2014
Report Date
April 15, 2014
Manufacturer
RESPIRONICS INC.,
Product Code
CBK
PMA / PMN Number
K093416
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE VENTILATOR WAS RETURNED TO THE THIRD PARTY SERVICE CENTER FOR EVALUATION AND THE CUSTOMER'S COMPLAINT WAS CONFIRMED. THE DEVICE'S ACTIVE EXHALATION CONTROL MODULE WAS REPLACED TO ADDRESS THE ISSUE.

Description of Event or Problem · 1

THE MANUFACTURER RECEIVED INFORMATION FROM A THIRD PARTY SERVICE CENTER ALLEGING A VENTILATOR FAILED A STEP DURING TESTING. THE DEVICE WAS NOT IN PATIENT USE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
275549 TRILOGY 200 VENTILATOR, CONTINUOUS, FACILITY USE CBK RESPIRONICS INC., 1032800

Patients

Seq Age Sex Outcome Treatment
1