COBAS INTEGRA 800
Report
- Report Number
- 1823260-2014-04739
- Event Type
- Malfunction
- Date Received
- June 30, 2014
- Date of Event
- May 30, 2014
- Report Date
- August 13, 2014
- Manufacturer
- ROCHE DIAGNOSTICS
- Product Code
- JJE
- PMA / PMN Number
- K951595
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
THIS EVENT OCCURRED IN (B)(6).
CORRECT RESULTS WERE RECEIVED UPON REPEAT TESTING AFTER THE SAMPLES WERE REMIXED, THE INVESTIGATION DETERMINED THE ISSUE WAS CAUSED BY SAMPLE SEDIMENTATION DUE TO EITHER A LONG WAIT ON THE ROTOR BEFORE SAMPLING OR DUE TO FASTER SEDIMENTATION OF SOME SAMPLES DUE TO A PATHOLOGICAL REASON. IT WAS RECOMMEND TO RUN THE SAMPLES IN SMALLER BATCHES AND TO AVOID A LONG WAIT ON THE ROTOR BEFORE PIPETTING. THE PROVIDED CALIBRATION AND QC DATA WAS ACCEPTABLE, A GENERAL ISSUE WITH THE INSTRUMENT OR REAGENT COULD BE EXCLUDED.
THE CUSTOMER RECEIVED QUESTIONABLE TINA-QUANT HEMOGLOBIN A1C GEN.2 (HBA1C) RESULTS FOR TWO PATIENT SAMPLES. PATIENT 1 ORIGINAL RESULT WAS 0.042 AND THE REPEAT RESULT WAS 0.061. PATIENT 2 ORIGINAL RESULT WAS 0.049 AND THE REPEAT RESULT WAS 0.065. NO UNIT OF MEASURE WAS PROVIDED. THE ORIGINAL RESULT WAS REPORTED OUTSIDE THE LAB. THE PATIENTS WERE NOT ADVERSELY AFFECTED. THE HBA1C REAGENT LOT NUMBER WAS 69089201. THE EXPIRATION DATE WAS REQUESTED, BUT WAS NOT PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 380069 | COBAS INTEGRA 800 | CLINICAL CHEMISTRY ANALYZER | JJE | ROCHE DIAGNOSTICS | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |