FDA Adverse Event Malfunction Summary report: N

RIATA ACTIVE FIXATION

MDR report key: 3901366 · Received June 30, 2014

Report

Report Number
2938836-2014-12619
Event Type
Malfunction
Date Received
June 30, 2014
Date of Event
May 5, 2014
Manufacturer
ST. JUDE MEDICAL, INC., CRMD
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.NO COMPLAINT RECEIVED WITH RETURN OF DEVICE. FAILURE (EVENT) OBSERVED DURING ANALYSIS.EXTERNAL INSULATION ABRASION WAS NOTED AT 45.4-45.6CM FROM THE DISTAL TIP, CONSISTENT WITH LEAD FRICTION TO THE ICD CAN. THE ETFE COATING WAS INTACT AT THIS LOCATION. INTERNAL INSULATION ABRASION WAS NOTED AT 13.4-14.5CM AND 15.5-17.4CM FROM THE DISTAL TIP. THE ETFE COATING WAS DAMAGED DURING THE SURGICAL PROCEDURE AT THESE LOCATIONS.(B)(4).

Description of Event or Problem · 1

THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
380938 RIATA ACTIVE FIXATION DEFIBRILLATION LEAD NVY ST. JUDE MEDICAL, INC., CRMD 1581/60 NA

Patients

Seq Age Sex Outcome Treatment
1