FDA Adverse Event Malfunction Summary report: N

RIATA ACTIVE FIXATION

MDR report key: 3901336 · Received June 30, 2014

Report

Report Number
2938836-2014-12721
Event Type
Malfunction
Date Received
June 30, 2014
Date of Event
April 17, 2014
Manufacturer
ST. JUDE MEDICAL, INC., CRMD
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
Z0457
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. EVALUATION DESCRIPTION: EXTERNAL INSULATION ABRASION WAS NOTED AT 8.0-8.2CM FROM THE CONNECTOR PIN, CONSISTENT WITH LEAD FRICTION TO THE ICD CAN. THE RV CONDUCTOR WAS EXPOSED AT THIS LOCATION. EXTERNAL AND INTERNAL INSULATION ABRASION WAS NOTED AT 15.0-15.7CM FROM THE DISTAL TIP. EXTERNALIZED CONDUCTORS DUE TO INTERNAL INSULATION ABRASION WERE NOTED AT 8.5-14.5CM FROM THE DISTAL TIP. THE ETFE COATING WAS INTACT AT THIS LOCATION.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING DEVICE CHANGE OUT FOR ERI, EXTERNALIZED CONDUCTORS WERE OBSERVED. NO ELECTRICAL ANOMALIES WERE NOTED. THE LEAD WAS EXPLANTED AND REPLACED. PATIENT WAS FINE BEFORE AND AFTER THE PROCEDURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
379942 RIATA ACTIVE FIXATION DEFIBRILLATION LEAD NVY ST. JUDE MEDICAL, INC., CRMD 1582/65 NA