FDA Adverse Event Malfunction Summary report: N

RIATA ACTIVE FIXATION

MDR report key: 3901260 · Received June 30, 2014

Report

Report Number
2938836-2014-12641
Event Type
Malfunction
Date Received
June 30, 2014
Date of Event
May 20, 2014
Manufacturer
ST. JUDE MEDICAL, INC., CRMD
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
Z0457
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. NO COMPLAINT RECEIVED WITH RETURN OF DEVICE. FAILURE (EVENT) OBSERVED DURING ANALYSIS. A PARTIAL LEAD WITH THE DISTAL TIP MEASURING 46.4CM WAS RETURNED FOR ANALYSIS. INTERNAL INSULATION ABRASION WAS NOTED AT 9.8-9.9CM FROM THE DISTAL TIP. EXTERNALIZED CONDUCTORS DUE TO INTERNAL INSULATION ABRASION WERE NOTED AT 9.8-12.6CM FROM THE DISTAL TIP. THE ETFE COATING WAS INTACT AT THESE LOCATIONS. RELIABILITY LABORATORY TECHNICIAN ST. JUDE MEDICAL CRMD RELIABILITY LABORATORY ST. JUDE MEDICAL CRMD RELIABILITY LABORATORY

Description of Event or Problem · 1

THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
380259 RIATA ACTIVE FIXATION DEFIBRILLATION LEAD NVY ST. JUDE MEDICAL, INC., CRMD 1580/65 NA

Patients

Seq Age Sex Outcome Treatment
1