RIATA ST OPTIM ACTIVE FIXATION
Report
- Report Number
- 2938836-2014-12668
- Event Type
- Injury
- Date Received
- June 30, 2014
- Date of Event
- May 7, 2014
- Manufacturer
- ST. JUDE MEDICAL, INC., CRMD
- Product Code
- NVY
- PMA / PMN Number
- P950022
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MS, US
- Reporter Occupation
- PHYSICIAN
Narratives
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.
A PARTIAL LEAD WAS RETURNED IN TWO SEGMENTS FOR ANALYSIS. THE DAMAGE FOUND WAS SUSTAINED DURING THE SURGICAL PROCEDURE. THE PORTIONS OF THE LEAD THAT WERE RETURNED WERE OTHERWISE NORMAL. WITHOUT RETURN OF THE ENTIRE LEAD, A COMPLETE ANALYSIS COULD NOT BE PERFORMED.
IT WAS REPORTED THAT THE PATIENT PRESENTED FOR A ROUTINE FOLLOW-UP. A STORED EGM SHOWED NOISE. THE NOISE COULD NOT BE REPRODUCED WITH ISOMETRIC TESTING. NO NOISE WAS SEEN AT A SUBSEQUENT FOLLOW-UP VISIT. MONITORING WILL CONTINUE. THERE WERE NO ADVERSE EFFECTS TO THE PATIENT.
NEW INFORMATION RECEIVED NOTES HIGH, OUT OF RANGE PACING LEAD IMPEDANCE. THE LEAD WAS EXPLANTED.
NEW INFORMATION RECEIVED NOTES THAT HIGH, OUT OF RANGE HIGH VOLTAGE LEAD IMPEDANCE WAS OBSERVED. THE PATIENT WAS ASYMPTOMATIC. LEAD REVISION IS PLANNED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 380799 | RIATA ST OPTIM ACTIVE FIXATION | DEFIBRILLATION LEAD | NVY | ST. JUDE MEDICAL, INC., CRMD | 7021/65 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 62 YR | Required Intervention | 1207-36, (B)(4) |