FDA Adverse Event Injury Summary report: N

DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR

MDR report key: 3901236 · Received June 30, 2014

Report

Report Number
2938836-2014-12612
Event Type
Injury
Date Received
June 30, 2014
Date of Event
April 15, 2014
Manufacturer
ST. JUDE MEDICAL, INC., CRMD
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. (B)(4). EVALUATION DESCRIPTION: A FRACTURE OF THE INNER COIL WAS FOUND AT THE CONNECTOR REGION CONSISTENT WITH FATIGUE. THIS FRACTURE IS CONSISTENT WITH THE OBSERVATION FROM THE FIELD OF NOISE AND LEAD IMPEDANCE MEASUREMENT ANOMALIES.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT RECEIVED INAPPROPRIATE THERAPY DUE TO NOISE. OUT OF RANGE AUTOCAPTURE MEASUREMENTS AND INCREASED PACING LEAD IMPEDANCE WERE ALSO NOTED. THE LEAD WAS EXPLANTED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
381124 DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR DEFIBRILLATION LEAD NVY ST. JUDE MEDICAL, INC., CRMD 7120Q/52 NA

Patients

Seq Age Sex Outcome Treatment
1 79 YR Required Intervention 2088TC/46, CAT024324