RIATA ACTIVE FIXATION
Report
- Report Number
- 2938836-2014-12616
- Event Type
- Malfunction
- Date Received
- June 30, 2014
- Date of Event
- April 25, 2014
- Manufacturer
- ST. JUDE MEDICAL, INC., CRMD
- Product Code
- NVY
- PMA / PMN Number
- P950022
- Removal / Correction Number
- Z0457
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. EVALUATION DESCRIPTION: NO COMPLAINT RECEIVED WITH RETURN OF DEVICE. FAILURE (EVENT) OBSERVED DURING ANALYSIS. EXTERNAL INSULATION ABRASIONS WERE NOTED AT 12.0-13.3CM AND 18.5-19.5CM FROM THE CONNECTOR PIN, CONSISTENT WITH LEAD FRICTION TO THE ICD CAN. EXTERNALIZED CONDUCTORS DUE TO INTERNAL INSULATION ABRASION WERE NOTED AT 9.5-12.8CM FROM THE CONNECTOR PIN. INTERNAL INSULATION ABRASIONS WERE NOTED AT 7.4-8.7CM, 7.5-7.7, 14.4-15.7CM, 15.3-16.0CM, AND 27.0-28.0CM FROM THE DISTAL TIP. THE ETFE COATING WAS INTACT AT THESE LOCATIONS. EXTERNAL INSULATION ABRASIONS WERE NOTED AT 12.5-12.7CM AND 17.8-19.4CM FROM THE CONNECTOR PIN, CONSISTENT WITH LEAD FRICTION TO THE ICD CAN. THE ETFE COATING WAS ABRADED AT 18.3-18.6CM FROM THE CONNECTOR PIN. THE CONDUCTOR WAS FRACTURED AND MELTED AT THIS LOCATION.
THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS.