FDA Adverse Event
Injury
Summary report: N
FORTIFY VR, DF-4 CONNECTOR
MDR report key: 3901133
·
Received June 30, 2014
Report
- Report Number
- 2938836-2014-12610
- Event Type
- Injury
- Date Received
- June 30, 2014
- Date of Event
- March 22, 2014
- Manufacturer
- ST. JUDE MEDICAL, INC., CRMD
- Product Code
- LWS
- PMA / PMN Number
- P910023
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PR, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. DEVICE EVALUATION ANTICIPATED, BUT NOT YET BEGUN.
Additional Manufacturer Narrative · 1
THE REPORTED PREMATURE BATTERY DEPLETION WAS CONFIRMED IN THE LABORATORY. THE DEVICE WAS TESTED ON THE BENCH AND AN ANOMALOUS COMPONENT ON THE INTEGRATED CIRCUIT WAS FOUND TO BE THE CAUSE OF THE PREMATURE BATTERY DEPLETION.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE DEVICE WAS EXPLANTED AND REPLACED DUE TO SUSPECTED PREMATURE BATTERY DEPLETION. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 379615 | FORTIFY VR, DF-4 CONNECTOR | IMPLANTABLE CARDIOVERTER DEFIBRILLATOR | LWS | ST. JUDE MEDICAL, INC., CRMD | CD1231-40Q | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 66 YR | Required Intervention |