UNIFY CRT-D
Report
- Report Number
- 2938836-2014-12661
- Event Type
- Injury
- Date Received
- June 30, 2014
- Date of Event
- April 22, 2014
- Manufacturer
- ST. JUDE MEDICAL, INC., CRMD
- Product Code
- NIK
- PMA / PMN Number
- P030054
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- PHYSICIAN
Narratives
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. DEVICE EVALUATION ANTICIPATED, BUT NOT YET BEGUN.
THE REPORTED FIELD EVENT OF AN INABILITY TO COMMUNICATE WITH THE DEVICE WAS CONFIRMED IN THE LABORATORY AND WAS FOUND TO BE CAUSED BY PREMATURE BATTERY DEPLETION. BASED ON ALL AVAILABLE PARAMETER AND USAGE INFORMATION, DEVICE LONGEVITY WAS FOUND TO BE BELOW THE EXPECTED LIMITS. WITH AN EXTERNAL POWER SUPPLY ATTACHED, THE DEVICE FUNCTIONED NORMALLY. NO HIGH CURRENT WAS DETECTED DURING TESTING AND THE POWER CONSUMPTION WAS NORMAL. THE ORIGINAL BATTERY WAS RETURNED TO THE VENDOR FOR FURTHER EVALUATION AND ANALYSIS COULD NOT CONCLUSIVELY DETERMINE A ROOT CAUSE FOR THE PREMATURE BATTERY DEPLETION.
IT WAS REPORTED THAT THE SYMPTOMATIC PATIENT PRESENTED IN THE EMERGENCY ROOM WITH SUSPECTED RAPID BATTERY DEPLETION. THE DEVICE WAS UNABLE TO BE INTERROGATED. THE DEVICE WAS EXPLANTED AND REPLACED. THE PATIENT TOLERATED THE PROCEDURE WELL WITH NO COMPLICATIONS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 378948 | UNIFY CRT-D | IMPLANTABLE CARDIOVERTER DEFIBRILLATOR | NIK | ST. JUDE MEDICAL, INC., CRMD | CD3231-40 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 71 YR | Required Intervention | 1582/65, RJ11272 |