FDA Adverse Event Injury Summary report: N

CURRENT RF DR

MDR report key: 3901026 · Received June 30, 2014

Report

Report Number
2938836-2014-12681
Event Type
Injury
Date Received
June 30, 2014
Date of Event
April 15, 2014
Manufacturer
ST. JUDE MEDICAL, INC., CRMD
Product Code
LWS
PMA / PMN Number
P910023
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MS, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE REPORTED FIELD EVENT OF AN INABILITY TO COMMUNICATE WITH THE DEVICE WAS CONFIRMED IN THE LABORATORY AND WAS DUE TO PREMATURE BATTERY DEPLETION. DEVICE LEVEL TESTING WITH AN EXTERNAL POWER SUPPLY INDICATED NORMAL ELECTRICAL PERFORMANCE. THE ORIGINAL BATTERY WAS RETURNED TO THE VENDOR FOR FURTHER EVALUATION AND ANALYSIS COULD NOT CONCLUSIVELY DETERMINE A ROOT CAUSE FOR THE PREMATURE BATTERY DEPLETION.

Additional Manufacturer Narrative · 1

DEVICE EVALUATION ANTICIPATED, BUT NOT YET BEGUN.

Description of Event or Problem · 1

IT WAS REPORTED THAT AN ASYMPTOMATIC PATIENT PRESENTED TO THE CLINIC WITH A DEVICE THAT FAILED TO INTERROGATE. AFTER MULTIPLE ATTEMPTS, TELEMETRY COULD NOT BE ESTABLISHED AND IT WAS SUSPECTED THE DEVICE WENT PREMATURELY INTO END OF SERVICE. THE DEVICE WAS EXPLANTED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
380594 CURRENT RF DR IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWS ST. JUDE MEDICAL, INC., CRMD 2207-36 NA

Patients

Seq Age Sex Outcome Treatment
1 73 YR Required Intervention