FENESTRATED BIPOLAR FORCEPSINSTRUMENT

MDR report key: 3900994 · Received June 30, 2014

Report

Report Number
2955842-2014-03978
Event Type
Malfunction
Date Received
June 30, 2014
Date of Event
May 7, 2014
Report Date
June 4, 2014
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, FAILURE ANALYSIS INVESTIGATION FOUND THE TIPS OF THE INSTRUMENTS GRIPS WERE BENT. ONE GRIP WAS BENT, CAUSING A SIDE TO SIDE MISALIGNMENT OF THE GRIPS. THERE WAS A .094 OFFSET AT THE TIPS. THE BENT GRIP DID NOT EXHIBIT SEPARATION OF THE YAW PULLEY AND PULLEY COVER; HOWEVER, THE LIKELY CAUSE OF BENDING IS OVERLOADING AT THE TIP. FAILURE ANALYSIS CONCLUDED THE DAMAGE MAY BE DUE TO MISHANDLING / MISUSE. AN ADDITIONAL FINDING WAS VARIOUS SCRATCHES ON THE DISTAL END OF THE MAIN TUBE SHOWING LIGHT MATERIAL REMOVAL. THE SCRATCHES WERE .086 - .126 IN LENGTH AND NOT AXIALLY ALIGNED WITH THE TUBE. NO OTHER DAMAGE WAS FOUND.

Description of Event or Problem · 1

IT WAS REPORTED THAT PRIOR TO STARTING A DA VINCI SURGICAL PROCEDURE, THE TIP WAS BENT ON THE FENESTRATED BIPOLAR FORCEPS INSTRUMENT. THERE WERE NO REPORTS OF FRAGMENTS FALLING INTO A PATIENT. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
379989 FENESTRATED BIPOLAR FORCEPSINSTRUMENT ENDOSCOPIC ELECTROSURGICAL INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420205-05 M10120920 905

Patients

Seq Age Sex Outcome Treatment
1