FDA Adverse Event Malfunction Summary report: N

COULTER LH 750 HEMATOLOGY ANALYZER

MDR report key: 3900973 · Received June 30, 2014

Report

Report Number
1061932-2014-01490
Event Type
Malfunction
Date Received
June 30, 2014
Date of Event
June 2, 2014
Report Date
June 2, 2014
Manufacturer
BECKMAN COULTER
Product Code
GKZ
PMA / PMN Number
K011342
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
LA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A BECKMAN COULTER (BEC) FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO ASSESS THE INSTRUMENT. THE FSE CONFIRMED THE LEAK FROM A HOLE IN THE YELLOW STRIPED TUBING THROUGH VALVE VL12A. THE FSE REPLACED THE TUBING TO RESOLVE THE LEAK ISSUE. THE FSE ALSO REPLACED ALL THE OTHER TUBING THROUGH VALVE VL12A, TUBING THROUGH CONNECTING Y FITTING, AND MIX CHECK VALVES AS PREVENTATIVE MAINTENANCE. THE FSE VERIFIED THE INSTRUMENT AND THE INSTRUMENT WAS RETURNED TO NORMAL OPERATION. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER REPORTED APPROXIMATELY 4 MLS OF FLUID LEAKED FROM THE COULTER LH 750 HEMATOLOGY ANALYZER ONTO THE COUNTER DURING STARTUP. THE LEAK WAS NOT CONTAINED WITHIN THE INSTRUMENT. THE OPERATOR WAS WEARING A LAB COAT AND GLOVES AT THE TIME OF THE INCIDENT. THERE WAS NO EXPOSURE TO OPEN WOUNDS OR MUCOUS MEMBRANES. NO ERRONEOUS PATIENT RESULTS WERE GENERATED IN CONNECTION WITH THIS EVENT. THERE WAS NO DEATH, INJURY, OR AFFECT TO USER OR TO PATIENT TREATMENT ATTRIBUTED TO THIS EVENT.